Thermogravimetric analyzers (TGA) measure weight change (loss or gain) and the rate of weight change as a function of temperature, time, and atmosphere.
TGA 3000 is an automated instrument which determines Moisture content, Ash content, Volatile content, Loss on Ignition (LOI) and Fixed Carbon content in a wide range of Organic, Inorganic and Synthetic materials.
Thermogravimetric analysis replaces the traditional analytical techniques that are slow, labor intensive and involve several steps with multiple laboratory equipment such us muffle furnaces, ovens and balances.
The TGA 3000 with integrated balance combines drying, ashing and weighing processes. This improves the efficiency, precision and provides high sample throughput.
The TGA 3000 is applicable in various industries and applications, including coal, coke, ores, cement, lime, foodstuffs, feeds and many more.
Furnace Temperature Range
|
Ambient to 1100 °C
|
Temperature Control Precision
|
±2% (or) ±2deg C
|
Accuracy
|
±2deg C of tolerance limit
|
Ramp Rate
|
10deg C /minute to 50deg C /minute
|
Balance
|
Integrated balance
|
Resolution
|
0.0001g (0.1mg)
|
Readability
|
0.0001g (0.1mg)
|
Sample Size
|
up to 5 grams (Higher range can be offered on request)
|
Number of Samples
|
19 Samples +1 Reference
|
Number of Carousels
|
|
Two (one for Crucibles and the other for Crucible lids)
|
|
Carousel Material
|
The carousels are made of special material which are not susceptible to warping under high temperature stress
|
Weighing Precision
|
|
0.02% RSD (on inert Samples)
|
|
Gas Connection
|
|
Oxygen & Nitrogen, Compressed Air
|
|
Power Supply
|
|
230V AC ± 10%; 50/60Hz/15A
|
TGA 3000 is not available for sale in USA.